• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented to the hospital for a scheduled implant procedure.During the procedure, the left ventricular (lv) lead had separation failure.The lv lead was removed and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of stylet could not be removed was confirmed.A complete lead with stuck stylet was returned in one piece.The ptfe coating of the stuck stylet was stripped and was found bunched up with the inner coil distal to the connector pin.The connector pin with the cap and crimp sleeve were found pulled out from the connector assembly stretching the inner coil consistent with damage due to excessive forces applied while attempting to remove the stylet from the lead.The cause of the reported events was isolated to the bunching of the stylet ptfe coating inside the inner coil at the connector region that prevented the removal of the stylet and excessive forces resulted in the connector pin with the cap and crimp sleeve to be pulled out through the connector assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18423229
MDR Text Key331793134
Report Number2017865-2023-95714
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/01/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000133147
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-