Model Number 37612 |
Device Problems
Failure to Deliver Energy (1211); Electromagnetic Compatibility Problem (2927)
|
Patient Problem
Ambulation Difficulties (2544)
|
Event Date 12/09/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the recharger is charged every friday.An unusual sound was heard during charging.It was later discovered that a metal tube of ointment was in the pocket near the stimulator.The body became immobile the next saturday so the stimulator was charged again, but the orange light came on.After a while, the light returned to green, so the stimulator was charged; however, the body remained immobile.The stimulator was charged twice yesterday, but the body is still immobile.When the patient was brought to juntendo hospital, it was told that stimulation was "paused." when the physician program was used, the body movements returned to normal and was given a dopamine infusion and went home.Additional information was received from the manufacturer representative (rep) that the power was off, probably due to some kind of electromagnetic interference.The patient came to the hospital and turned the stimulation back on and resolved the problem.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|