• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM; OLYMPUS HYSTERORESECTOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH TELESCOPE, 30°, 4 MM; OLYMPUS HYSTERORESECTOSCOPE Back to Search Results
Model Number A22002A
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation, and the customers allegation was confirmed.The evaluation found the following reportable issues: the non-olympus outer tube and laser markings where on the light guide post with a missing direction pin.The following non-reportable malfunctions were found during the device evaluation: the fiber was broken, there were dents on the frame, there was debris under the eyepiece window with minor scratches on the funnel, the optical fiber system had a foggy image.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that the telescope, 30°, 4 mm has a piece missing in the arbitrator hole.The issue was found during preparation for use.There were no reports of patient harm.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Corrected fields: d8.Additional information added to field h3, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, the cause of this malfunction was identified to be third party repair of the device.Three attempts were performed to obtain additional information, but no response was received from the customer.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 30°, 4 MM
Type of Device
OLYMPUS HYSTERORESECTOSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18427697
MDR Text Key331944326
Report Number9610773-2024-00050
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761020855
UDI-Public04042761020855
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K897003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22002A
Device Lot Number663646
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/11/2023
Initial Date FDA Received01/02/2024
Supplement Dates Manufacturer Received04/22/2024
Supplement Dates FDA Received04/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-