C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Poor Quality Image (1408); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/11/2023 |
Event Type
malfunction
|
Event Description
|
It was reported image is not clear.
|
|
Manufacturer Narrative
|
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and will evaluate.Results are expected soon.
|
|
Manufacturer Narrative
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was returned to the service facility for evaluation.During evaluation, the reported issue of unclear image was unconfirmed.Per service technician, there were no imaging issues found.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.
|
|
Event Description
|
It was reported image is not clear.
|
|