• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. WEB SL; INTRASACCULAR FLOW DISRUPTION DEVICE Back to Search Results
Model Number MV-WB050321
Device Problem Difficult or Delayed Separation (4044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  Injury  
Manufacturer Narrative
Investigation findings: the device remains implanted in the patient and not available for return to the manufacturer for evaluation.Visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): please refer to the japanese ifu for precautions, warnings, and further information.The following is taken from the english version: potential complications potential complications include but are not limited to the following: hematoma at the site of entry, aneurysm rupture, emboli, vessel perforation, parent artery occlusion, hemorrhage, ischemia, vasospasm, clot formation, device migration or misplacement, premature or difficult device detachment, non-detachment, incomplete aneurysm filling, revascularization, post-embolization syndrome, and neurological deficits including stroke and death.Warnings and precautions the web aneurysm embolization system is intended for single use only.The detachment control device is intended to be used for one patient.Do not re-sterilize and/or reuse the device.Reuse and/or re-sterilization can increase risk of infection, cause a pyrogenic response or other life-threatening complications.Reuse and/or re-sterilization can degrade product performance, leading to device malfunction.Dispose of all devices in accordance with applicable hospital, administrative and/or local government policy.Advance and retract the device slowly.Do not advance the delivery device with excessive force.Determine the cause of any unusual resistance.Remove the device if excessive friction is noted and check for damage.Do not rotate the delivery device during or after delivery of the embolization device.Rotating the device may result in damage or premature detachment.The embolization device cannot be detached with any other power source other than a microvention inc.Detachment control device.Ensure that at least two detachment control devices are available before initiating an embolization procedure.Procedure detachment of the device 31.The detachment control device is pre-loaded with batteries and will activate when the delivery device is properly connected.32.Verify that the rhv is firmly locked around the delivery device before attaching the detachment control device to ensure that the embolization device does not move during the connection process.33.Ensure that the delivery device gold connectors are clean and free from blood or contrast.If necessary, wipe the connectors with sterile water and dry before connecting.34.Insert the proximal end of the delivery device into the detachment control device.When the delivery device is properly connected, the light will flash green, and an intermittent tone will be heard.35.Verify the embolization device position before pressing the detachment button.36.Push the detachment button.During firing, the light should be solid green, and the beep should be continuous.37.Verify detachment by first loosening the rhv valve, then pulling back slowly on the delivery device and verifying that there is not embolization device movement.If the embolization device does not detach, push the detachment button again.If the device is still not detached, obtain a new detachment control device and attempt detachment up to two additional times.If it does not detach, remove the delivery device.38.Verify the position of the embolization device angiographically through the guide catheter.39.Prior to removing the microcatheter from the treatment site, place an appropriately sized guidewire completely through the microcatheter lumen to ensure that no part of the embolization device remains within the microcatheter.Investigation conclusion: the physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.
 
Event Description
It was reported that during an embolization procedure, the web device was deployed and confirmed by conebeam ct to be positioned properly within the target aneurysm.The first attempt to detach the web with the detachment controller was unsuccessful.A second attempt to detach was made by advancing the tip of the catheter close to the embolization device and pulling the delivery device, but detachment failed.Electrical continuity was normal as the light on the controller flashed green in both attempts at detachment.The web was then advanced to the tip of the microcatheter and was retrieved in to the microcatheter.Upon manipulation of the delivery device and advancing the web device, the web was found to be detached in the aneurysm unintentionally, resulting in a portion of the embolization device protruding slightly into the parent vessel.Angiography revealed that the embolization device was implanted without any issues.Therefore, the patient was decided to be placed under observation and the procedure was completed, and dapt will be performed for an extended period.There was no health damage to the patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Type of Device
INTRASACCULAR FLOW DISRUPTION DEVICE
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key18437181
MDR Text Key331866751
Report Number2032493-2024-00027
Device Sequence Number1
Product Code OPR
UDI-Device Identifier04987892122309
UDI-Public(01)04987892122309(11)220907(17)270831(10)0000253811
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P170032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMV-WB050321
Device Lot Number0000253811
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
VIA 21 MICROCATHETER
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-