• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TRANSPAC® IV MONITORING KIT, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH DEVICE,; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. TRANSPAC® IV MONITORING KIT, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH DEVICE,; TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR Back to Search Results
Catalog Number 886-42584-05
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2023
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of leakage was confirmed.No samples or videos were returned for evaluation.However, an image was provided by the customer showing a leak between the pvc tubing and the drip chamber.Without the return of the reported sample, a probable cause could not be determined.The lot history was reviewed and there were no nonconformities which would have contributed to the reported complaint.
 
Event Description
The event involved a transpac® iv monitoring kit, disposable transducer, 03 ml squeeze flush device, macrodrip where it was reported that there was leakage on connector which may result in patient discomfort or an uncontrolled condition and may have to extend their hospital stay.The solution/medication involved was normal saline.There is unknown patient involvement and unknown patient harm.There was no chemo drug involved and the other information was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPAC® IV MONITORING KIT, DISPOSABLE TRANSDUCER, 03 ML SQUEEZE FLUSH DEVICE,
Type of Device
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18437401
MDR Text Key331916134
Report Number9617594-2024-00006
Device Sequence Number1
Product Code DRS
UDI-Device Identifier00840619068615
UDI-Public(01)00840619068615(17)260301(10)13596341
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K061573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number886-42584-05
Device Lot Number13596341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NORMAL SALINE, MFR UNK
-
-