Model Number LXMJ37L
Device Problems Mechanical Problem (1384); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found the tip of the jaw was broken.Piece on the broken jaw was missing.Functional testing noted that the jaws could not open or close.The knife blade could not advance due to the damage to the jaws.It was reported that the device did not cut sufficiently and the knife blade could not advance.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: avoid grasping objects, such as staples, clips, or encapsulated sutures in the jaws of the instrument.D10 concomitant product: vlft10gen, vlft10gen ft series energy platformx1 (serial#(b)(6); medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during total laparoscopic hysterectomy procedure, the device did not cut sufficiently.The knife blade could not advance.Replaced with another device and it worked fine.There was no patient injury.Medtronic's initial evaluation of the incident device found jaws cannot open.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18438157
MDR Text Key331828586
Report Number1717344-2024-00025
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10884521830936
UDI-Public10884521830936
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K223158  
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLXMJ37L
Device Catalogue NumberLXMJ37L
Device Lot Number30460172X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received01/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PLEASE SEE NOTE ON H10
Patient Age51 YR
Patient SexFemale