• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problems Moisture or Humidity Problem (2986); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) customer called on behalf of the bedside rn, for assistance with several error codes on console.The balloon used was an axillary inserted sensation plus 50cc that had been in for 11 days.Customer had reported that they had been dealing with condensation in the helium line for several hours and the console alarmed "error code 4" then system failure.It was reported they switched out the console.There was no patient harm reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
A getinge field service engineer evaluated the unit and replaced screw kit, cardiosave preventive mainten, assy,helium reservior, pressure transducer, 30 psia, 4"cable, drive manifold assembly, pneumatic module assy, rohs, compressor, fitting,muffer to reservoir, o-ring, buna, 1 muffler,1/8 npt, muffler <100 micron, screw kit, cardiosave pneumatic assembly and passed all functional and safety tests.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18439733
MDR Text Key331939459
Report Number2249723-2024-00027
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/23/2023
Initial Date FDA Received01/03/2024
Supplement Dates Manufacturer Received04/22/2024
Supplement Dates FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SENSATION PLUS 50CC.
Patient Age60 YR
Patient SexMale
Patient Weight105 KG
-
-