• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX BLUE LINE ULTRA SUCTIONAID TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. PORTEX BLUE LINE ULTRA SUCTIONAID TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/860/080
Device Problem Inflation Problem (1310)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Manufacturer Narrative
Month and year of event have been provided, day is unknown.D4: lot number, udi number, expiration date, and h4: manufacture date are unknown; no information has been provided to date.E2 - incorrect due to system limitations.E2 correct response: no, non-healthcare professional.H3 - other: device has not been returned to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that there was water accumulation in the cuff.The event occurred in late (b)(6) 2023.This occurred during patient use.No patient harm/adverse event reported.
 
Manufacturer Narrative
One device was received for investigation.Per visual inspection, the sample doesn´t show any signs of damage and no water inside the pilot balloon and cuff observed.The device was then functionally tested.No leaks were found.The reported complaint is not confirmed.A review of the device history records could not be completed as no lot number was provided by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE ULTRA SUCTIONAID TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18440317
MDR Text Key332613706
Report Number9617604-2024-00005
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K030570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number100/860/080
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/01/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-