Lot Number 6002330
Device Problem Material Twisted/Bent (2981)
Patient Problem Diabetic Ketoacidosis (2364)
Event Type  Injury  
Event Description
Unomedical reference number (b)(4).Event occurred in the (b)(6).It was reported that the patient faced a bent cannula which led to high blood glucose level as she was not getting insulin.Therefore, they tried to treat it with correction bolus, but on (b)(6) 2023, the patient went to the emergency room and was subsequently hospitalized due to high blood glucose level.Her highest blood glucose level was 1200 mg/dl.Moreover, the infusion set had been used for one day.Further, she was transferred to the intensive care unit.During hospitalization, the patient received fluids of saline, insulin, and unspecified (drug name unknown) medication as corrective treatment which resolved the issue.On (b)(6) 2023, the patient was released from the hospital with no permanent damage.Unomedical do not see bent/kinking as being related to human factors, but rather as a training issue including correct choices of insertion sites and infusion sets and cannula length.Furthermore, the soft cannula is a flexible material that during use and upon removal can bend slightly.No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSOFT XC
Type of Device
UNO INSET I 60/9 GREY TCAP 10PK INT
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key18441815
MDR Text Key331864499
Report Number3003442380-2024-00107
Device Sequence Number1
Product Code FPA
UDI-Device Identifier05705244016651
UDI-Public05705244016651
Combination Product (y/n)Y
PMA/PMN Number
K032854
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/04/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Lot Number6002330
Date Manufacturer Received12/28/2023
Patient Sequence Number1
-
-