Model Number 10-85-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland has received a report of error/alarm shaft angel encoder fault.There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 double head pump the incident occurred in north kansas city, missouri.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The problem was traced back to a defective shaft encoder and was solved by part replacement.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
|
|
Manufacturer Narrative
|
The involved roller pump was manufactured in (b)(6) 2009 and according to the analysis of the complaints database no further events have been reported in the past.Based on the above facts and taking into account the investigation performed for similar events, the reported fault can be solely due to a defective shaft angle encoder.Taking into account the manufacturing year of the unit, it cannot be ruled out that wearing of the part that can be originated by the specific use at customer's facility may have contributed to the event.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Event Description
|
See initial report.
|
|