Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that on 2023-12-29, the site left the system unplugged overnight.The representative told them to keep the system plugged into wall power, hoping the batteries would recover.The system appeared to be functioning so the account decided to use it for a case.The system performed as expected throughout the entirety of the case.When returning the system back to its storage space after thecase, the image acquisition system (ias) died during transport.The representative checked the system again today 2024-1-1 but the image acquisition system (ias) was dead. there was no patient involvement.
|
H3) onsite functional and visual examination was performed by a manufacturer representative.The batteries were replaced and the issue was resolved.The system passed a system checkout and was returned to an operational condition.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: kit svc o2 bi71000162r batt tray 2pk rfb, x2 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|