Model Number 0031588667
Device Problems Use of Device Problem (1670); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  Injury  
Manufacturer Narrative
Block h6: imdrf device code a150101 captures the reportable event of stent failure to expand.
 
Event Description
It was reported to boston scientific corporation that an axios stent and electrocautery enhanced delivery system was to be implanted during a gastrojujenostomy procedure performed on (b)(6) 2023.During the procedure, the axios stent could not expand well in the gastric end.A balloon device was used to dilate the stent, however, the axios stent shrink in the gastric end under fluoroscopy.The stent remains implanted, and the procedure was completed.There were no known patient complications reported as a result of this event.The patient is currently under observation to check the stent function.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Type of Device
PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18445628
MDR Text Key331918423
Report Number3005099803-2023-07022
Device Sequence Number1
Product Code PCU
Combination Product (y/n)N
Reporter Country CodeTW
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/09/2024
Device Model Number0031588667
Device Catalogue Number5356
Device Lot Number0031588667
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2023
Initial Date FDA Received01/04/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-