• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134805
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter for which biosense webster¿s product analysis lab (pal) identified a hole in the pebax.Initially it was reported that there was a force sensor error.There was no patient consequence.The biosense webster, inc.Product analysis lab received the device for evaluation and per the evaluation completion on 08-dec-2023, there was a hole and reddish material in the pebax.This event was originally considered non-reportable, however, bwi became aware of a hole in the pebax on 08-dec-2023 and have assessed this returned condition as reportable.
 
Manufacturer Narrative
The device evaluation was completed on 08-dec-2023.The smart touch bidirectional sf device was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed a hole and reddish material in the pebax.The device was connected to carto 3 system and was visualized and recognized correctly; however, error 106 appeared on the system due to an open circuit in the tip area.The blood found inside the pebax area may contribute to error 106.A manufacturing record evaluation was performed and no internal actions was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18452116
MDR Text Key332731589
Report Number2029046-2024-00071
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134805
Device Lot Number31085464L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2023
Date Manufacturer Received12/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-