Catalog Number D134805 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter for which biosense webster¿s product analysis lab (pal) identified a hole in the pebax.Initially it was reported that there was a force sensor error.There was no patient consequence.The biosense webster, inc.Product analysis lab received the device for evaluation and per the evaluation completion on 08-dec-2023, there was a hole and reddish material in the pebax.This event was originally considered non-reportable, however, bwi became aware of a hole in the pebax on 08-dec-2023 and have assessed this returned condition as reportable.
|
|
Manufacturer Narrative
|
The device evaluation was completed on 08-dec-2023.The smart touch bidirectional sf device was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed a hole and reddish material in the pebax.The device was connected to carto 3 system and was visualized and recognized correctly; however, error 106 appeared on the system due to an open circuit in the tip area.The blood found inside the pebax area may contribute to error 106.A manufacturing record evaluation was performed and no internal actions was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.This product issue will be addressed through bwi's quality system.Manufacturer's reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|