BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404231
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem Insufficient Information (4580)
Event Date 12/21/2023
Event Type  Injury  
Event Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced floppy glans syndrome due to the cylinders not being long enough for his anatomy.A revision surgery was performed to explant and replaced the cylinders with a larger one.No additional patient complications were reported.
 
Manufacturer Narrative
There was no device available for analysis; therefore, no physical or visual analysis of the product could be performed.The cylinders being too short for the patient anatomy could have caused or contributed to the reported patient symptom of floppy glans, which is a known risk associated with implant of these devices as indicated in the instructions for use (ifu).Based on the information available, a conclusion code of unintended use error caused or contributed to event was assigned to this investigation as the event was determined to be unintentional or inadvertent interaction.
 
Event Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced floppy glans syndrome due to the cylinders not being long enough for his anatomy.A revision surgery was performed to explant and replaced the cylinders with a larger one.No additional patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18455272
MDR Text Key332254790
Report Number2124215-2023-75002
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953003443
UDI-Public00878953003443
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/25/2013
Device Model Number72404231
Device Catalogue Number72404231
Device Lot Number739094003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received01/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RESERVOIR: UPN: 72404155 / LOT: 743110002; RESERVOIR: UPN: 72404155 / LOT: 743110002
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age58 YR
-
-