• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL AIRLIFE¿ MISTY MAX WITH PEDIATRIC AEROSOL MASK AND OXYGEN TUBING; NEBULIZER (DIRECT PATIENT INTERFACE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL AIRLIFE¿ MISTY MAX WITH PEDIATRIC AEROSOL MASK AND OXYGEN TUBING; NEBULIZER (DIRECT PATIENT INTERFACE) Back to Search Results
Model Number MASK AEROSOL PED W/MISTY & 7FT TBG 50/CS
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2023
Event Type  Injury  
Event Description
The customer reported that the female end of the ck1001 - mask aerosol ped w/misty & 7ft tbg 50/cs was not able to easily be attached to an oxygen tree, and the flange prevented a few of their compressor units from being used.Staff have been working around this issue by using additional tubing; however, this is leading to waste, which they would prefer to avoid.The patient's impact or harm is currently unknown.
 
Manufacturer Narrative
H3: 81 other - at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Manufacturer Narrative
Based on the investigation and since the customer did send photos of the fg ck1001 with lot number 0004250616 for the investigation, we require the physical sample and/or pictures as evidence to perform a better investigation to confirm the reported defect and determine the cause.In addition, the device history record is reviewed as part of the investigation; however, since the lot number was not provided, the review was not possible.Additionally, a sampling was performed by quality personnel on component female u-connector part number 65-4198c (used for the manufacture of fg ck1001) in inventory by dimensionally inspecting the cavity where it is assembled, finding all inspected samples within specifications.Based on the above, the reported defect was not confirmed, and the root cause was not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRLIFE¿ MISTY MAX WITH PEDIATRIC AEROSOL MASK AND OXYGEN TUBING
Type of Device
NEBULIZER (DIRECT PATIENT INTERFACE)
Manufacturer (Section D)
VYAIRE MEDICAL
26125 north riverwoods blvd
mettawa IL 60045
Manufacturer (Section G)
PRODUCTOS UROLOGOS DE MEXICO S.A. DE C.V.
cerrada vía de la producción
parque undustrial mexicali iii
mexicali 21397
MX   21397
Manufacturer Contact
sandra valencia
510 technology drive
irvine, CA 92618
2402760001
MDR Report Key18455415
MDR Text Key332203418
Report Number8030673-2024-00369
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K790892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMASK AEROSOL PED W/MISTY & 7FT TBG 50/CS
Device Catalogue NumberCK1001
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-