• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL G4 STEERABLE GUIDING CATHETER (CE); CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL G4 STEERABLE GUIDING CATHETER (CE); CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0702
Device Problems Break (1069); Difficult or Delayed Positioning (1157); Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Event Description
It was reported during preparation the steerable guide catheter (sgc), the +/- knob was turned, but the device did not straighten.It was noted a cable break was suspected.Therefore, the sgc was not used and was replaced.There was no clinically significant delay in the procedure.No additional information provided.
 
Manufacturer Narrative
The device has been received.However, investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
The returned device analysis was unable to confirm the reported cable break and difficult or delayed positioning associated with difficult to curve and straighten.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.All information was investigated and based on the information provided and the returned device analysis (unable to confirm the reported issues), a cause for the reported difficult to curve and straighten cannot be determined.The reported broken cable was due to the user speculation.There is no indication of a product issue with respect to manufacture, design or labeling.H6: code 1158 removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER (CE)
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18456065
MDR Text Key332207684
Report Number2135147-2024-00110
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Catalogue NumberSGC0702
Device Lot Number30201R1053
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2023
Initial Date FDA Received01/05/2024
Supplement Dates Manufacturer Received01/12/2024
Supplement Dates FDA Received01/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-