Catalog Number SGC0702 |
Device Problems
Break (1069); Difficult or Delayed Positioning (1157); Positioning Failure (1158)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2023 |
Event Type
malfunction
|
Event Description
|
It was reported during preparation the steerable guide catheter (sgc), the +/- knob was turned, but the device did not straighten.It was noted a cable break was suspected.Therefore, the sgc was not used and was replaced.There was no clinically significant delay in the procedure.No additional information provided.
|
|
Manufacturer Narrative
|
The device has been received.However, investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Manufacturer Narrative
|
The returned device analysis was unable to confirm the reported cable break and difficult or delayed positioning associated with difficult to curve and straighten.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.All information was investigated and based on the information provided and the returned device analysis (unable to confirm the reported issues), a cause for the reported difficult to curve and straighten cannot be determined.The reported broken cable was due to the user speculation.There is no indication of a product issue with respect to manufacture, design or labeling.H6: code 1158 removed.
|
|