MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Defective Alarm (1014)
Patient Problem Insufficient Information (4580)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
It was reported that not the device was not passing the overtemperature alarm test.Patient involvement is unknown.
 
Manufacturer Narrative
B3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received.Per visual inspection, the water tank was cover cracked and leaked from the bottom two screws holes, quick connect was corroded, and faded line cord.Outdated printed circuit board (pcb) and power switch.Per functional testing, the device did not pass the over temperature alarm test confirming the customer complaint.The root cause was the device needed to be recalibrated.It was not known what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18457984
MDR Text Key332505621
Report Number3012307300-2024-00103
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer12/19/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received01/05/2024
Supplement Dates Manufacturer Received02/07/2024
Supplement Dates FDA Received03/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-