• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACON LABORATORIES, INC. FLOWFLEX COVID-19 ANTIGEN HOME TEST; ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACON LABORATORIES, INC. FLOWFLEX COVID-19 ANTIGEN HOME TEST; ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP Back to Search Results
Lot Number COV3090007
Device Problem Unable to Obtain Readings (1516)
Patient Problem Insufficient Information (4580)
Event Date 01/03/2024
Event Type  malfunction  
Manufacturer Narrative
The current complaint is pending investigation from our contract manufacturer, and we will provide follow up mdr upon investigation completion such as review of batch records or testing of retention samples.Any additional information received by acon may be provided to fda in a follow up mdr.
 
Event Description
Invalid result (s).Invalid results.The customer stated that the kits they purchased from walmart did not produce a control (c) or test (t) line.
 
Manufacturer Narrative
Final product manufacture and qc record for cov3090007.No abnormal issue was found in manufacturing process, technical testing and quality control inspection, and the manufacturing process is complied with dmr.The test results of retention samples from cov3090007 can meet the qc criteria.We have not found the complaint issue from the retained cassettes.The complaint is not verified.In this follow-up report, the following information has been updated from the initial report upon completion of the internal investigation.
 
Event Description
Invalid results.The customer stated that the kits they purchased from walmart did not produce a control (c) or test (t) line.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOWFLEX COVID-19 ANTIGEN HOME TEST
Type of Device
ANTIGEN DIAGNOSTIC FOR SARS-COV-2, PRODUCT CODE: QKP
Manufacturer (Section D)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer (Section G)
ACON LABORATORIES, INC.
5850 oberlin drive #340
san diego CA 92121
Manufacturer Contact
qiyi xie
MDR Report Key18460339
MDR Text Key332733243
Report Number2531491-2024-00002
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
EUA210494
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Lot NumberCOV3090007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/03/2024
Initial Date FDA Received01/05/2024
Supplement Dates Manufacturer Received01/03/2024
Supplement Dates FDA Received01/19/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-