Model Number 1192
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Implant Pain (4561)
Event Date 12/11/2023
Event Type  Injury  
Event Description
Related manufacturer reference number: 1627487-2024-00143.It was reported that the patient lost about 60 pounds and the lead and anchor were protruding from the midline incision site.There was no skin erosion.It was noted that the anchor was too shallow and causing irritation.No falls or trauma were reported.Surgical intervention was undertaken wherein the lead and anchor were explanted.Effective therapy restored post-op.
 
Manufacturer Narrative
Section b3: date of event is estimated.
 
Manufacturer Narrative
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18460966
MDR Text Key332257871
Report Number1627487-2024-00144
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model Number1192
Device Lot Number5457578
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2023
Initial Date FDA Received01/05/2024
Supplement Dates Manufacturer Received01/20/2024
Supplement Dates FDA Received01/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS IPG (1); SCS PERCUTANEOUS LEAD (1)
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight50 KG
-
-