Model Number MI1000 MED-EL CONCERT |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
|
Patient Problems
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/15/2023 |
Event Type
malfunction
|
Event Description
|
The user visited the clinic with bloody ear discharge and no more hearing benefit.Upon examination, the doctor observed that the ear was wet and noticed the extrusion of the electrode lead, which was visible in the ear canal.As a result, the doctor recommended re-implantation.
|
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The user visited the clinic with bloody ear discharge and no more hearing benefit with the device.The problem started 10-12 days prior (mid december 2023).The user has been reimplanted with a new device.
|
|
Manufacturer Narrative
|
Conclusions: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received device was explanted due to an extrusion of the electrode lead into the external ear canal.Reportedly the main reason for the extrusion was due to a device unrelated enlarged cholesteatoma and granulation.In addition, the recipient suffered from an ear infection.A review of the device_s sterilization records shows that the device has been subject to a valid sterilization process, thus no available information points to the implant being the source of infection.The user has history of ear infections prior to implantation.This is a final report.
|
|
Search Alerts/Recalls
|