• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problems Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
Patient Problems Headache (1880); Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 12/08/2023
Event Type  Injury  
Event Description
An "unspecified" error message was reported with the adc device and customer was unable to obtain readings.As a result, customer experienced a loss of consciousness, headache and was unable to self-treat.The customer had contact with a healthcare professional (hcp) who administered glucose tablets and orange juice for treatment of a hypoglycemia diagnosis.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18462980
MDR Text Key332255976
Report Number2954323-2024-00684
Device Sequence Number1
Product Code QLG
UDI-Device Identifier00357599800000
UDI-Public00357599800000
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/30/2024
Device Model Number71992-01
Was Device Available for Evaluation? No
Date Manufacturer Received12/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-