LIVANOVA DEUTSCHLAND GMBH S5 CONSOLE FOR 3 PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 48-30-00 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland has received a report that, during a procedure, the s5 console displayed error code when trying to pump.There is no report of any patient involvement.
|
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the s5 console.The incident occurred in usa.Field service was dispatched to customer site, ran the system and found the motor controller failed error code.Motor control board and power end-stage board were replaced and when the console was restarted, it returned the same error.The old motor control board was reinstalled, a new power end-stage was installed and ran system with no issues.The device has been returned to customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
Complaints database analysis revealed no similar event since unit installation in 2011.Based on the above, the root cause of the event was traced back to a defective hmf board.Failure of electronic boards can be related to multiple and not deterministic factors such as exposure to heat, dust and moisture, accidental impact (drops and falls), and power overloads/surges but also to variability of micro subcomponents.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Event Description
|
See initial report.
|
|