• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ALINITY HQ PROCESSING MODULE; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ALINITY HQ PROCESSING MODULE; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 09P68-01
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2023
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
 
Event Description
The customer observed falsely decreased hemoglobin (hgb) patient results on the alinity hq processing module.The samples retested higher on a different module.One example was provided.(b)(6)2023 female patient who had ectopic pregnancy alinity hq hgb result of 73 g/l was reported out of the lab and questioned by the ward.Repeat testing generated a hgb result of 123 g/l.A new sample was drawn and the hgb result was 128 g/l.No impact to patient management was reported.
 
Manufacturer Narrative
The field service representative (fsr) inspected the instrument.The fsr found heavy crystallization around hgb reagent supply valves (125 &126), and replaced valves and fittings.The hgb cradle 1 was also noted to be crystalized; however, it was not determined to have impacted hgb results.Trending review did not identify any trends.Additionally, labeling was reviewed and adequately addressed the issue under review.Based on the investigation, no deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY HQ PROCESSING MODULE
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
ABBOTT LABORATORIES
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT LABORATORIES
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden N39 E-932
GM   N39 E932
6122582960
MDR Report Key18465278
MDR Text Key332374242
Report Number2919069-2024-00003
Device Sequence Number1
Product Code GRZ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K220031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09P68-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/08/2024
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received02/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
-
-