This complaint was received via user report and has been reported to fda.A valid serial number has not been provided.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.Clinical data was reviewed and confirmed that freestyle libre sensors continue to be safe, effective, and perform as intended in the field.The available tripped trend reports were reviewed for libre sensor and perception code for the last year.The review identified a tripped trend for this perception code.This tripped trend was addressed in the tracking, trending review meeting, and investigated.The investigation concluded that the tripped trend was not correlated with a product nonconformance.Trends are regularly monitored and investigated when exceeding established thresholds to evaluate for causes associated with the product.If the product is returned, a physical investigation will be performed and a follow-up report submitted.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
Abbott diabetes care received a medwatch report which reported the following information: the customer indicated the adc device was "defective" as it did not communicate with the reader.There was no report of an adverse event or third-party intervention due to the reported issue.Adc customer service attempted to contact the reporter 3 times to gain additional details regarding this event; however, all follow up attempts were unsuccessful.Based on the information provided, there was no report of serious injury associated with this event.
|