It was reported to boston scientific corporation that a cre pro gi wireguided dilatation balloon was used during a procedure performed on (b)(6) 2023.During the procedure, the balloon burst.The procedure was completed with another cre pro gi wireguided dilatation balloon.There were no patient complications reported as a result of this event.Additional information received on january 14, 2024.The type of procedure performed was lower gastrointestinal dilatation.The customer reported that no pieces of the balloon detached inside the patient.
|
Block h6: imdrf device code a0402 captures the reportable event of a balloon burst.Block h10: investigation results: the returned cre pro gi wireguided dilatation balloon was analyzed, and a visual examination found that the balloon and catheter of the device had no damages.Functional analysis was performed, and the balloon was inflated without a problem.However, the balloon would not hold pressure due to a pinhole in the balloon (distal section), located approximately 44 mm from the tip.Microscopic inspection found a pinhole located approximately at 44 mm from the tip of the device.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon burst was not confirmed.The pinhole problem found could have been interpreted by the customer as the reported event of a balloon burst.The pinhole found is likely to have occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during or previous the procedure, could have caused the problem found on the distal section of the balloon.Therefore, the most probable root cause is an adverse event related to procedure.Block h11: additional information: block b5 (describe event or problem).
|