BOSTON SCIENTIFIC CORPORATION MOMENTUM CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G125
Device Problems Failure to Capture (1081); High impedance (1291); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2023
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this system with a cardiac resynchronization therapy defibrillator (crt-d) and a non-bsc left ventricular (lv) lead triggered an alert for lv automatic threshold test.It appeared that the test had not been able to be run for various days, nonetheless, when the crt-d was analyzed, it was no longer in suspension mode.The impedance values had increased as well and several months ago the behavior had been unstable.Also, failure to capture was observed for the lv lead.Before the high, out of range pacing impedances the presenting electrogram appeared different and the lead was capturing.A lead fracture was suggested, and the health care professional agreed.The crt-d and the lv lead remain in service.No symptoms or adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this system with a cardiac resynchronization therapy defibrillator (crt-d) and a non-bsc left ventricular (lv) lead triggered an alert for lv automatic threshold test.It appeared that the test had not been able to be run for various days, nonetheless, when the crt-d was analyzed, it was no longer in suspension mode.The impedance values had increased as well and several months ago the behavior had been unstable.Also, failure to capture was observed for the lv lead.Before the high, out of range pacing impedances the presenting electrogram appeared different and the lead was capturing.A lead fracture was suggested, and the health care professional agreed.The crt-d and the lv lead remain in service.No symptoms or adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOMENTUM CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18472844
MDR Text Key332900642
Report Number2124215-2024-00747
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589126
UDI-Public00802526589126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/23/2023
Device Model NumberG125
Device Catalogue NumberG125
Device Lot Number147683
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-