Model Number 0600-0783
Device Problems Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2023
Event Type  malfunction  
Event Description
The customer reported that during a patient procedure, using a glidescope core 15-inch monitor, the picture on the screen froze when attempting to intubate the patient.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Manufacturer Narrative
The device return is anticipated, however; at the time of the report the device has not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
 
Manufacturer Narrative
The subject glidescope core smart cable used during the reported event was returned to verathon for evaluation.A verathon technical service representative evaluated the returned smart cable and was able to confirm the reported cable failure.Upon visual inspection, it was identified that the smart cable's hdmi connector was damaged by a blade retention pin.Upon completion of verathon's device evaluation, the smart cable was scrapped due to the customer already being provided a replacement and there being no repairs available for the device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Manufacturer Narrative
Following the submission of verathon's initial mdr, the customer notified verathon that they had isolated the issue to the glidescope core smart cable rather than the glidescope core 15-inch monitor.It was reported that the smart cable had a damaged prong in the hdmi connector, which was shown in a photo provided by the customer.Additionally, the customer reported that the incident occurred in (b)(6) 2023 but due to miscommunication within their facility, it was not reported to verathon until (b)(6) 2023.The event date and device details have been updated accordingly within this supplemental report.A replacement glidescope core smart cable was provided to the customer and the damaged glidescope core smart cable was requested to be returned to verathon for evaluation.The device return is anticipated, however; at the time of the report the device has not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key18476614
MDR Text Key332990129
Report Number9615393-2024-00009
Device Sequence Number1
Product Code CCW
UDI-Device Identifier00879123006639
UDI-Public010087912300663911200505
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0600-0783
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/21/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-