H3, h6: the reported device was received for evaluation.A visual inspection revealed that the device was not returned in any original packaging.One side of the suture capture has been partially disconnected from the upper jaw.No other visual deficiencies were found.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include (1) excessive force, (2) tissue thickness, (3) and/or damage or debris on the device tip between passes.No containment or corrective actions are recommended at this time.
|