Trackwise id#: (b)(4).Updated sections: b-4, g-4, g-7, h-2, h-3, h-6, h-10 the device was returned to the factory for evaluation on (b)(6) 2023.An investigation was conducted on (b)(6) 2024.A visual inspection was conducted.Signs of clinical use and evidence of blood was observed.The heater wire was observed to be intact with no visual defects observed.The gray silicone insulation on the hot jaw was observed the to be intact with no visual defects observed.The gray silicone insulation on the cold jaw was observed to be peeled away from the base of the cold jaw to the closest part to the tip of the cold jaw.The detached silicone insulation was not returned for evaluation.Based on the returned condition of the device as well as the evaluation results, the reported failure "peeled; delaminated; jaw" was confirmed.The lot # 3000355798 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|