ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY CRT-D; NO MATCH Back to Search Results
Model Number CD3231-40
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for implant procedure.During procedure it was noted that the setscrew fell off the pacemaker.The procedure was completed using a different pacemaker.The patient was in stable condition.
 
Manufacturer Narrative
The reported field event of set screw anomaly was confirmed.Final analysis found that the right ventricular setscrew was missing, and the rv septum was damaged upon received.This is consistent with the reported field event of setscrew fell out during operation.The rv septum was damaged during operation could cause the setscrew fell out when it was over loosened.The device was above elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18480775
MDR Text Key332724038
Report Number2017865-2024-00752
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734504362
UDI-Public05414734504362
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCD3231-40
Device Lot NumberA000144715
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received01/09/2024
Supplement Dates Manufacturer Received01/22/2024
Supplement Dates FDA Received01/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-