PRO MED INSTRUMENTS GMBH DORO® QR3 SKULL CLAMP Back to Search Results
Model Number 1001.001
Device Problem Device Slipped (1584)
Patient Problem Laceration(s) (1946)
Event Date 12/08/2023
Event Type  Injury  
Manufacturer Narrative
As the product was in specification and did not show any deviations that could cause the reported incident, we suspect, that maybe the pinning technique has been not optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event Description
On (b)(6) a customer informed us that one of our products was involved in a case (tumor resection, prone position) in which the treated patient sustained a laceration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® QR3 SKULL CLAMP
Type of Device
DORO® QR3 SKULL CLAMP
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
nicholas preissler
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key18481690
MDR Text Key332491535
Report Number3003923584-2023-00031
Device Sequence Number1
Product Code HBL
UDI-Device Identifier04250435506196
UDI-Public04250435506196
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K191740
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1001.001
Device Catalogue Number1001.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2023
Initial Date Manufacturer Received 12/11/2023
Initial Date FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-