MAUDE Adverse Event Report: MEDTRONIC IRELAND HAWKONE 7FR; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND HAWKONE 7FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number H1-LS-INT
Device Problems Difficult to Remove (1528); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
Physician was attempting to use a hawkone ls atherectomy device along with a 7fr sheath and an unknown guidewire during treatment of a calcified plaque lesion in patient.A 6mm embolic protection device was used.The ifu (instruction for use) was followed during preparation, procedure, post-procedure.Resistance was not felt during advancement.Damaged components during procedure are reported.The flush port on catheter handle was not damaged.A bulge in the cutter and difficulty when trying to remove through sheath are reported.The damaged component did not need to be retrieved from patient.Procedure was completed as normal.No patient injury reported.When the device was returned to the manufacturing facility, it was noted that that the tecothane wall at the bulge was not intact.
 
Manufacturer Narrative
Product analysis the device was returned with a bulge on the nosecone approx.21mm from the cutter window.The cutter is stuck in the biologics and could not be retracted or advanced, the tecothane wall at the bulge is not intact, medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAWKONE 7FR
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18482341
MDR Text Key332494279
Report Number9612164-2024-00183
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K141801
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberH1-LS-INT
Device Lot Number0011213094
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-