• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 1758SI14
Device Problems Decrease in Pressure (1490); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the foley catheter was naturally removed.The patient was kept in the operating room and discovered to be missing in the ward.
 
Manufacturer Narrative
The reported event is confirmed - cause unknown.Received two (2) photo samples.Both photo samples showcase an overview of all-silicone foley catheter with sample port connector.Received one (1) used all-silicone foley catheter with sample port connector without original packaging.Visual inspection noted the balloon has tear measuring 0.1460".Further inspection noted the balloon had no missing pieces.This is out of specification which states, "balloon must not be torn".A potential root cause for this failure could be pinhole in balloon or cuff.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.The instructions for use were found adequate and states the following: "[warnings] 1.Method for use: (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients: ·patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients: patients with known allergy to silver coated catheter." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the foley catheter was naturally removed.The patient was kept in the operating room and discovered to be missing in the ward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® I.C. ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18483558
MDR Text Key333097945
Report Number1018233-2024-00043
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K984136
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1758SI14
Device Lot NumberMYHV1882
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-