Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.A visual evaluation showed the device was returned in original packaging with the batch number of the complaint on the label.There is no visible defect.A functional evaluation showed pulling the lever will close and lock the jaw.Pulling the lever and trigger simultaneously will not deploy the suture passer needle.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the failure include excessive force or torsion during use, use of the needle as a lever, misplaced force on the device actuator.No containment or corrective actions are recommended at this time.
|