• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Catalog Number NGE-017115
Device Problems Break (1069); Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.G4: pma/510k # ¿ exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
As reported, an ngage nitinol stone extractor was placed through an unspecified scope and when they attempted to open the basket it would not open.The extractor was removed from the scope and two ends of the basket were found broken off.Nothing had to be removed from the patient.Another same type device was used to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.There were no adverse effects on the patient due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key18488007
MDR Text Key332901052
Report Number1820334-2024-00055
Device Sequence Number1
Product Code FFL
UDI-Device Identifier10827002482941
UDI-Public(01)10827002482941(17)260718(10)15564544
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNGE-017115
Device Lot Number15564544
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-