• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q190V
Device Problem Microbial Contamination of Device (2303)
Patient Problem Unspecified Infection (1930)
Event Type  Injury  
Manufacturer Narrative
This report has been submitted by the importer under this mdr report number 2429304-2024-00019.Additional information revealed that the faclity use sterile water and inject it into the instrument channel of the scope.The water is collected in a container at the distal tip of the scope.They use a sterile swab w iping the distal tip of the scope as well as the elevator.The swab is then placed in the same sample of water from the instrument channel.Only one sample is sent off for evaluation.Due to the asset listing error, the customer has not been able to try anything with the scope.It is currently quarantined and not in use.This scope was cultured tjf-q190v and came back fine with no positive growth.The device referenced in this report was not returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The customer reported to olympus that during routine microbiological testing, the evis exera iii duodenovideoscope tested positive for an unexpected contamination.The testing was performed after a therapeutic endoscopic retrograde cholangiopancreatography procedure.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.Additional information from the olympus employee reported that upon follow-up at the site and discussing the positive culture report with the customer, it was discovered the facility had two potential patients post procedure infections.Both instances the patient had an ercp- endoscopic retrograde cholangiopancreatography procedure before the infection and a culture positive scope was used in the procedure.The customer has not disclosed the date the scopes were cultured, and the type/growth of microbiological organism tested.The facility is reported to be inquiring further details from the infection prevention control.2 similar occurrences have been reported by the customer for patients having post-procedure infections at their facility with the patient identifiers: 1) patient identifier # (b)(6) evis exera iii duodenovideoscope tjf-q190v, sn (b)(6).2) patient identifier # (b)(6) evis exera iii duodenovideoscope tjf-q190v sn -unk (this report).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18490052
MDR Text Key332603780
Report Number9610595-2024-00676
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170452024
UDI-Public04953170452024
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K220587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q190V
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/11/2023
Initial Date FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-