Catalog Number 50000000 |
Device Problems
Circuit Failure (1089); Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
|
Event Description
|
It was reported that biomed had the arctic sun device with alarm 78 (non recoverable system error) when powered on, explained that was a bad thermistor and they would need a new tank harness.
|
|
Manufacturer Narrative
|
The reported issue was confirmed.The root cause of the reported issue was failed t4 thermistor.The dhr is not required as this is event not an out-of-box failure and therefore is not manufacturing related.The complaint or reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
|
|
Event Description
|
It was reported that biomed had the arctic sun device with alarm 78 (non recoverable system error) when powered on, explained that was a bad thermistor and they would need a new tank harness.
|
|