MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000; ARCTIC SUN DEVICE Back to Search Results
Catalog Number 50000000
Device Problems Circuit Failure (1089); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that biomed had the arctic sun device with alarm 78 (non recoverable system error) when powered on, explained that was a bad thermistor and they would need a new tank harness.
 
Manufacturer Narrative
The reported issue was confirmed.The root cause of the reported issue was failed t4 thermistor.The dhr is not required as this is event not an out-of-box failure and therefore is not manufacturing related.The complaint or reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the device was not returned.
 
Event Description
It was reported that biomed had the arctic sun device with alarm 78 (non recoverable system error) when powered on, explained that was a bad thermistor and they would need a new tank harness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN® 5000
Type of Device
ARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18493472
MDR Text Key332670806
Report Number1018233-2024-00084
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number50000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-