Model Number 4FC12
Device Problems Break (1069); Peeled/Delaminated (1454); Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2023
Event Type  malfunction  
Event Description
It was reported that during a cryo ablation procedure, the dilator could not be fixed to the sheath.The sheath was replaced which resolved the issue.It was then reported that the sheath was unable to be aspirated with the balloon catheter.The balloon catheter was further inserted without resolution.The balloon catheter was then replaced which resolved the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot number 0011872141 was returned and analyzed.External visual inspection was performed and the handle, shaft and sideport were intact with no apparent issues.The dilator visual inspection identified that the dilator luer was delaminated.Functional testing was performed.The dilator was inserted into the sheath and was unable to lock the dilator luer to the sheath.During the microscope inspection, it was identified that the dilator luer was damaged.In conclusion, the dilator unable to lock into the sheath was confirmed through analysis and the sheath failed return inspection due to the dilator luer delamination and damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18494190
MDR Text Key332935217
Report Number3002648230-2024-00010
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00763000418052
UDI-Public00763000418052
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0011872141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AFAPRO28 BALLOON CATHETER
Patient Age55 YR
Patient SexMale
-
-