Product event summary: the 4fc12 sheath with lot number 0011872141 was returned and analyzed.External visual inspection was performed and the handle, shaft and sideport were intact with no apparent issues.The dilator visual inspection identified that the dilator luer was delaminated.Functional testing was performed.The dilator was inserted into the sheath and was unable to lock the dilator luer to the sheath.During the microscope inspection, it was identified that the dilator luer was damaged.In conclusion, the dilator unable to lock into the sheath was confirmed through analysis and the sheath failed return inspection due to the dilator luer delamination and damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|