Model Number 72081-01
Device Problem Low Readings (2460)
Patient Problems Hyperglycemia (1905); Vomiting (2144); Dizziness (2194)
Event Date 12/19/2023
Event Type  Injury  
Manufacturer Narrative
The reported product is not expected to be returned as reporter indicated the device was discarded.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The device history records (dhrs) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and sensor kit passed all tests prior to release.In the event that unanticipated product is received, a physical investigation will be performed per adc's established processes and procedures and a follow-up report will be submitted upon completion of investigation.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A low reading issue was reported with the adc device.A customer reported unspecified low sensor readings and experienced symptoms of vomiting and lightheadedness.As a result, the customer was unable to self-treat and presented to the local hospital.Upon arrival, the customer was seen by a healthcare professional (hcp) who drew blood and administered insulin (type/dose unknown) for treatment.There was no report of death or permanent injury associated with this event.A sensor reading of 95 mg/dl was compared to a built-in meter reading of 214 mg/dl, and the results when plotted on a parkes error grid and fell into the c zone showing the difference in values to be clinically significant.However, it is unknown when these readings were taken in relation to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 3
Type of Device
CONTINUOUS GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE LTD
range road
witney 12345 -700
UK  12345-7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18495089
MDR Text Key332683450
Report Number2954323-2024-01371
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K223435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number72081-01
Was Device Available for Evaluation? No
Date Manufacturer Received01/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-