Model Number 0998-00-3023-45
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/02/2024
Event Type  malfunction  
Manufacturer Narrative
Due to character restrictions in block e1 telephone number : (b)(6).A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during routine inspection, the cs300 intra-aortic balloon pump (iabp) could not be turned on, immediately the equipment was no longer used.There were no causalities reported.
 
Manufacturer Narrative
A getinge field service engineer evaluated unit.The device cannot be turned on, re-plug and unplug the motherboard cable and reinstall the motherboard.The fault disappears.The equipment has passed all performance and safety tests specified by the factory.The machine performs normally and is delivered to the customer for normal use.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, CHINESE, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18498135
MDR Text Key332919720
Report Number2249723-2024-00105
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108575
UDI-Public10607567108575
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-45
Device Catalogue Number0998-00-3023-45
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/02/2024
Initial Date FDA Received01/11/2024
Supplement Dates Manufacturer Received02/23/2024
Supplement Dates FDA Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-