This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition that the device displayed an e6(overheat warning) error code and ran in a locked position, identified during service and repair, was confirmed.The assignable root cause was determined to be traced to maintenance.Udi: (b)(4).
|
It was reported that the motor device had an undetermined malfunction.During in-house engineering evaluation, it was observed that the device displayed an e6(overheat warning) error code, had foreign substance/debris/cleaning/sterilization and ran in a locked position.It was further determined that the device failed pretest for motor thermistor assessment and safety assessment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|