It was reported during preventative maintenance, the device had an unstable motor speed, and the motor was corroded.There was no patient involvement.There are no adverse events associated with this malfunction.Due diligence is complete there is no further information available.
|
This event has been recorded, by zimmer biomet under (b)(4).Review of the most recent repair record determined, the motor was corroded and the speed was unstable.The motor was replaced and resolved, the reported issue.Review of the device history record identified, no deviations or anomalies, during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found, which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|