• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Gas/Air Leak (2946)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) had a helium leak.Customer reported a rapid loss of helium.
 
Event Description
It was reported that before use, the cardiosave intra-aortic balloon pump (iabp) had a helium leak.Customer reported a rapid loss of helium.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: b4, d9, g3, g6, h2, h3, h4, h6, h10.A getinge field service engineer (fse) was dispatched to investigate the issue.After the inspection, the fse found that the iabp failed the helium leak check by dropping more than 20psi in 5 minutes.The fse also found that when the iabp console was removed from the hospital cart, more than 3 psi was lost in 30 minutes from the internal tank.The fse replaced the helium reservoir (d997-00-0565).After installing the part, the helium leak was no longer present.The fse then performed a pm, full calibration, and tested the unit.The product passed all functional/safety testing and was returned to the customer.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18502961
MDR Text Key332904538
Report Number2249723-2024-00109
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-