RESPIRONICS, INC. DREAMSTATION AVAPS30 AAM; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number FRX1130S14
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2023
Event Type  malfunction  
Event Description
The manufacturer received information regarding a dreamstation avaps30 aam.The device was returned to a third-party service center for evaluation.There was no allegation of serious or permanent harm or injury.No medical intervention was required for the patient.The third-party service center visually inspected the device.During evaluation, corrosion was found on the power connector.In addition, corrosion on metallic surface and dust/dirt were also observed on the device.This incident has been deemed reportable due to the corrosion found on the power connector.The device was scrapped.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.
 
Manufacturer Narrative
H3 other text : device evaluated by third party service center.
 
Manufacturer Narrative
The manufacturer received information in relation to a dreamstation avaps unit.The device was returned to a third party service center.During visual inspection of the device, it was determined the power connector of the unit was corroded and the unit didn't power on.Corrosion on metallic surfaces and dust/dirt contamination were also found during device evaluation.Device was scrapped at customer's request.Analysis of past events has not indicated an adverse event has occurred due to corrosion.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.Additionally, the risk file indicates that corrosion will not substantially affect the performance of the device.This complaint is considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AVAPS30 AAM
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18503977
MDR Text Key332974091
Report Number2518422-2024-02230
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K102465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberFRX1130S14
Device Catalogue NumberFRX1130S14
Was Device Available for Evaluation? No
Date Manufacturer Received02/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/30/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-