Model Number 8300AB
Device Problem Device Stenosis (4066)
Patient Problem Heart Failure/Congestive Heart Failure (4446)
Event Date 06/28/2023
Event Type  Injury  
Manufacturer Narrative
The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potentials known and unknown patient-related contributing factors.Per technical summary 33069, rev a, structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.Devices implanted 5 years or greater with calcification, dehiscence, leaflet tears, thickened leaflet, pannus, or structural valve deterioration (stenosis, regurgitation) reported as the complaint do not require an engineering evaluation.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A capa/scar/pra is not required as there are no confirmed product or labeling non-conformances and no other triggers are met.The most likely cause is patient factors, including history of hyperlipidemia (hld) and coronary artery disease (cad).All pertinent information available to edwards lifesciences has been submitted.
 
Event Description
It was reported and learned through medical records that a patient with a 25mm 8300ab aortic valve in aortic position underwent a valve in valve procedure after an implant duration of five (5) years, three (3) months and 23 days due to severe aortic stenosis.The patient was presented with acute on chronic chf, nyha 111.Tavr procedure was completed with a non-edwards transcatheter valve and patient was discharged home on pod# 7 in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key18509771
MDR Text Key332880500
Report Number2015691-2024-00393
Device Sequence Number1
Product Code LWR
UDI-Device Identifier00690103194500
UDI-Public(01)00690103194500(17)210826
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/26/2021
Device Model Number8300AB
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Life Threatening;
Patient Age76 YR
Patient SexMale
-
-