Model Number PM3222
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Pocket Erosion (2013)
Event Date 01/03/2024
Event Type  Injury  
Manufacturer Narrative
The sterilization records were reviewed, and no evidence of abnormal sterilization cycle was found.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented with an open sore scab at the pocket site due to infection.The whole system was explanted, and the pacemaker was left exteriorly taped to the patient's body with a temporary right ventricular lead implanted through the right jugular vein.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURE RF
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18510115
MDR Text Key332868834
Report Number2017865-2024-01114
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734507905
UDI-Public05414734507905
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model NumberPM3222
Device Lot NumberA000128166
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUICKFLEX; TENDRIL
Patient Outcome(s) Required Intervention;
Patient Age91 YR
Patient SexFemale
-
-