• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND MICRA AV; LEADLESS PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND MICRA AV; LEADLESS PACEMAKER Back to Search Results
Model Number MC1AVR1
Device Problems Coagulation in Device or Device Ingredient (1096); Mechanical Problem (1384); Use of Device Problem (1670); Deformation Due to Compressive Stress (2889); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2023
Event Type  malfunction  
Manufacturer Narrative
Continuation of d10: this is a system report.The section d information is for the primary device, which was in use with the following: brand name [micra] product id [mc1avr1- delsys] (serial: (b)(6).4092-58 lead, implanted: (b)(6) 2004.4592-53 lead, implanted: (b)(6) 2004 w3dr01 ipg, implanted: (b)(6) 2023 product id 5076-52 lot# serial# (b)(6) implanted: (b)(6) 2023.Explanted: (b)(6) 2023-12-29 product id 5076-58 lot# serial#(b)(6) implanted: (b)(6) 2023.Explanted: (b)(6) 2023 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced pocket erosion and wound dehiscence approximately one month post implant of a right atrial (ra) lead and a right ventricular (rv) lead, which were connected to an existing implantable pulse generator (ipg).It was reported that the ipg system was exposed to the air and was contaminated but not necessarily infected.The ipg system was explanted and the pocket was debriefed, respected and closed to prevent any future infection.A leadless ipg was then attempted to be implanted in the rv septum.It was reported that intra-operative, the leadless ipg system was caught in previously capped leads.It was reported that there were possible tissue and blood clots between the device and the cup of the delivery system, and the ipg was unable to be retracted into the cup despite flushing.It was also reported that delivery system was bent.The leadless ipg system was attempted/not used and was replaced.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Correction: section b1 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced pocket erosion and wound dehiscence approximately one month post implant of a right atrial (ra) lead and a right ventricular (rv) lead, which were connected to an existing implantable pulse generator (ipg).It was reported that the ipg system was exposed to the air and was contaminated but not necessarily infected.The ipg system was explanted and the pocket was debriefed, respected and closed to prevent any future infection.A leadless ipg was then attempted to be implanted in the rv septum.It was reported that intra-operative, the leadless ipg system was caught in previously capped leads.It was reportedthat there were possible tissue and blood clots between the device and the cup of the delivery system, and the ipg was unable to be retracted into the cup despite flushing.It was also reported that delivery system was bent.The leadless ipg system was attempted/not used and was replaced.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA AV
Type of Device
LEADLESS PACEMAKER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18510757
MDR Text Key333539090
Report Number9612164-2024-00236
Device Sequence Number1
Product Code PNJ
UDI-Device Identifier00763000078621
UDI-Public00763000078621
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC1AVR1
Device Catalogue NumberMC1AVR1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2024
Initial Date FDA Received01/15/2024
Supplement Dates Manufacturer Received01/15/2024
Supplement Dates FDA Received01/15/2024
Date Device Manufactured09/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10...."
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age71 YR
Patient SexMale
-
-