• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR; DBS IPG Back to Search Results
Model Number 6662
Device Problem Migration (4003)
Patient Problem Movement Disorder (4412)
Event Date 01/03/2024
Event Type  Injury  
Manufacturer Narrative
B3-date of event is estimated.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Event Description
Related manufacturer reference number 1627487-2024-00292; related manufacturer reference number 1627487-2024-00293.It was reported that patient experience ineffective therapy.X-rays showed that the ipg flipped in the pocket and extension were twisted as well.As a result, surgical intervention was undertaken wherein the extensions and ipg were replaced and repositioned.Effective therapy was restored post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ 7 IMPLANTABLE PULSE GENERATOR
Type of Device
DBS IPG
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
heidi syndergaard
6901 preston road
plano, TX 75024
9723098000
MDR Report Key18511380
MDR Text Key332886768
Report Number1627487-2024-00291
Device Sequence Number1
Product Code MHY
UDI-Device Identifier05415067020260
UDI-Public05415067020260
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P140009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number6662
Device Lot NumberT00004988
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-