LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-50-00
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a s5 roller pump gave a motor control failure error message during procedure.There was no patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 system.The incident occurred in joliet, illinois.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
From follow-up communication, field service engineer responsible for the site clarified that no service activity has been performed since the customer did not request it, likely due to an isolated occurrence of the error.Error 442 (fault in motor controller: detection of speed set value different to actual value) can be caused by an occlusion too hard, wrong tubing, rotor blocked by foreign material or defective rotor bearing.Taking into account that the error has not come back, it can be excluded that the issue was related to defective bearings.A device service history review has been performed and identified that the unit was manufactured in 2018 and no other similar event has been reported, neither concerning trend has been identified.As per ifu, the user should check all cables, tubes, connectors and other accessories to make sure that they are connected correctly, are not leaking and are in perfect working order.Based on all the above, the most likely root causes of the reported event are the following:(i) occlusion too hard, (ii) wrong tubing and or/ (iii) rotor blocked by foreign material.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18516170
MDR Text Key333487265
Report Number9611109-2024-00039
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier04033817900900
UDI-Public(01)04033817900900(11)181217
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-50-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received01/16/2024
Supplement Dates Manufacturer Received02/20/2024
Supplement Dates FDA Received03/20/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/17/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-